Model Number FX2100X15B
Medical Device Problem Codes Circuit Failure (1089); Electrical /Electronic Property Problem (1198)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging a trilogy evo o2 ventilator experienced an unidentified device failure.There was no serious patient harm or injury that occurred or was reported.The trilogy evo o2 ventilator was returned and evaluated by a third-party service center.The service technician states that a ventilator inoperative error code relating to a user interface software version mismatch was found in the device's error log.The display printed circuit assembly (pca) board was replaced to resolve the issue.Additionally, a non-reportable motor controller shut down error code was also identified and resolve by replacing the system pca board.Also, a non-reportable corrupt software certificate error code was found in the device's error log and also resolved by the display pca replacement.The air inlet foam filter and particulate filter were replaced for maintenance.The device passed all final testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY EVO
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23303140
Report Number2518422-2025-111036
Device Sequence Number11958036
Product Code CBK
UDI-Device Identifier00606959061019
UDI-Public(01)00606959061019
Combination Product (Y/N)N
Initial Reporter CountryPL
PMA/510(K) Number
K181166
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberFX2100X15B
Device Catalogue NumberFX2100X15B
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2025
Initial Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-