A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown.
|
|
Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: - deflation: observed delamination of diaphragm valve through microscopic inspection assessed as adhesive failure.No further actions are required as it is not related to the manufacturing process.- capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.- crease / folding of implant: observed creases on device through visual inspection.No further actions are required as it is not related to the manufacturing process.- capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (wear abrasion) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: a5, a6, b5, d4, d9, e1, h3, h6.
|