• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN? EL ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN? EL ICD DR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number D143
Medical Device Problem Codes Defibrillation/Stimulation Problem (1573); Inaccurate Synchronization (1609)
Health Effect - Clinical Code Ventricular Fibrillation (2130)
Date of Event 09/22/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) presented with a rhythm that the physician determined was a slow ventricular tachycardia (vt).A burst of commanded anti-tachycardia pacing (atp) was delivered, followed by burst of device driven atp, resulting in the rhythm accelerating into ventricular fibrillation (vf).Then, this device successfully delivered a shock and converted the rhythm.Technical services (ts) was contacted and analyzed the episode data and noted the device appeared to be operating as programmed.This icd remains in service.No additional adverse patient effects were reported.
 
Event or Problem Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) presented with a rhythm that the physician determined was a slow ventricular tachycardia (vt).A burst of commanded anti-tachycardia pacing (atp) was delivered, followed by burst of device driven atp, resulting in the rhythm accelerating into ventricular fibrillation (vf).Then, this device successfully delivered a shock and converted the rhythm.Technical services (ts) was contacted and analyzed the episode data and noted the device appeared to be operating as programmed.This icd remains in service.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
The device remains in service; therefore, a technical analysis could not be performed.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.A risk review was completed and confirmed that the event of rhythm acceleration was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of known inherent risk of device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN? EL ICD DR
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23304281
Report Number2124215-2025-73501
Device Sequence Number18637078
Product Code LWS
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P960040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/05/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/12/2021
Device Model NumberD143
Device Catalogue NumberD143
Device Lot Number218543
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2025
Supplement Date Received by Manufacturer04/29/2026
Initial Report FDA Received Date10/15/2025
Supplement Report FDA Received Date05/05/2026
Date Device Manufactured08/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Other;
Patient Age67 YR
Patient SexMale
-
-