| Catalog Number |
367884 |
| Medical Device Problem Code |
Short Fill (1575)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® lithium heparinn (lh) 75 usp units blood collection tubes, an unspecified number of tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|
Additional Manufacturer Narrative
|
|
The following fields were updated due to additional information: e.1.Initial reporter addr 2: (b)(6).Investigation summary: bd had not received any samples or photos for investigation.A total of 100 retained samples were visually inspected, with the hemogard closure assembly correctly assembled and placed on the tubes.Additionally, a functional test was performed on 10 retained samples for low or no draw and all tubes drew within specification.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: underfill.No definitive root cause could be established for the reported defect.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® lithium heparinn (lh) 75 usp units blood collection tubes, an unspecified number of tubes underfilled.No adverse impact was reported regarding the patient or user.
|
| |
|
Search Alerts/Recalls
|