Model Number SK53
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 09/24/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was reported that after an initial bunion surgery on (b)(6) 2025, all hardware was removed during a revision surgery on (b)(6) 2025 due to loss of correction and the medial implant being displaced.The patient was non-compliant, fell down, and injured her foot.Upon removal of the hardware, it was confirmed the medial cuneiform was fractured.Additionally, it was noted that the patient had rheumatoid arthritis and poor bone quality.The surgeon chose not to replace the implants at that time.An update from the surgeon indicates an external rail was later inserted and the patient continued to be non-compliant.No deficiencies or malfunctions were alleged/found with any tmc device, and no other patient outcomes were reported as a result of this event.No devices were returned for evaluation.The device history records were reviewed, and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.Based on the available information, the most likely cause of what was reported is patient non-compliance and poor bone quality.No deficiencies or malfunctions have been alleged/found with any tmc device.The company will supplement this mdr as necessary and appropriate.Additional tmc devices explanted in the same removal surgery were reported in 3011623994-2025-00110.
 
Event or Problem Description
It was reported that after an initial bunion surgery on (b)(6) 2025, all hardware was removed during a revision surgery on (b)(6) 2025 due to loss of correction and the medial implant being displaced.The patient was non-compliant, fell down, and injured her foot.Upon removal of the hardware, it was confirmed the medial cuneiform was fractured.No deficiencies or malfunctions were alleged/found with any tmc device, and no other patient outcomes were reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPIPLASTY SPEEDPLATE
Common Device Name
STAPLE
Manufacturer (Section D)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer (Section G)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer Contact
uriza shums
100 palmetto park place
ponte vedra, FL 32081
9043735940
MDR Report Key23304697
Report Number3011623994-2025-00109
Device Sequence Number9980431
Product Code JDR
UDI-Device Identifier00810111222396
UDI-Public(01)00810111222396(17)290312(10)300531516
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K251135
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSK53
Device Lot Number300531516
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/24/2025
Initial Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
Patient SexFemale
-
-