| Catalog Number |
468-300 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
07/07/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade unknown.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture baker grade unknown and exchange from textured to smooth breast implant due to the patient¿s concern with the product.".This record is for the right side.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: h6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture baker grade unknown and exchange from textured to smooth breast implant due to the patient¿s concern with the product.".This record is for the right side.The device remains implanted.
|
| |
|
Search Alerts/Recalls
|