ABBOTT VASCULAR INC. TREK¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 1012272-15
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Visual inspection was performed on the returned device.The reported separation was confirmed.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the reported information and the observations from the returned analysis, the investigation determined that the reported separation appears to be related to operational context.Return device analysis notes that the hypotube fracture face was in oval shaped which can indicate the hypotube was kinked and then separated.There was no damage noted to the balloon dilatation catheter (bdc) during the inspection prior to use which suggests a product quality issue did not contribute to the reported difficulties.In this case, it is possible that the bdc was inadvertently mishandled during preparation and/or advancement such that the shaft became kinked and upon further manipulation it ultimately separated.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed in the posterior descending artery with heavy calcification.The 2.5x15mm trek rx balloon dilatation catheter (bdc) was advancing on the guide catheter without resistance when the hypotube snapped.There was no excessive force applied.The separated portions were simply pulled and removed from the guide catheter.The guide catheter remained in the anatomy and was used with another trek rx that completed the procedure.There was no adverse patient effect and there was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREK¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela UNK
CS   UNK
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key23306289
Report Number2024168-2025-11135
Device Sequence Number5135050
Product Code LOX
UDI-Device Identifier08717648138294
UDI-Public(01)08717648138294(17)280630(10)50730G1
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K180040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012272-15
Device Catalogue Number1012272-15
Device Lot Number50730G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/22/2025
Initial Report FDA Received Date10/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/30/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexMale
Patient Weight122 KG
-
-