• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY ASSURA VR ICD, US; Implantable cardioverter defibrillator (non-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) FORTIFY ASSURA VR ICD, US; Implantable cardioverter defibrillator (non-CRT) Back to Search Results
Model Number CD1357-40C
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/25/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The cause of infection could not be traced to the device.Further information was requested but not received.
 
Event or Problem Description
It was reported the patient presented to the hospital due to infection.The implantable cardioverter defibrillator (icd) and right ventricular lead were explanted.The patient status was unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FORTIFY ASSURA VR ICD, US
Common Device Name
Implantable cardioverter defibrillator (non-CRT)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key23307767
Report Number2017865-2025-1002785
Device Sequence Number12209537
Product Code LWS
UDI-Device Identifier05414734507943
UDI-Public(01)05414734507943(10)P000172594(17)250531
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model NumberCD1357-40C
Device Catalogue NumberCD1357-40C
Device Lot NumberP000172594
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/01/2025
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2025
Initial Report FDA Received Date10/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RIATA ACTIVE LEAD.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-