| Catalog Number |
FSENCOR01 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.D2b: (gei; knw).Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an upright vertical stereotactic procedure, the encor driver would not stop sampling after the user removed their foot off the encor foot pedal.The encor enspire needed to be switched off to stop the driver from sampling.After the enspire was switched back on, error code 1005 (driver button or contact stuck during initialization) was displayed.The enspire was switched off before removing the needle from the patient.There was patient involvement with no injury.It was identified that the encor enspire , the encor foot pedal and the encor driver were used together as a system during this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a dhr and manufacturing review was not required as the event was determined to be expected, and the investigation did not identify any manufacturing and/or service-related issues.Investigation summary: the foot switch serial number 04998 was received at the bard medical south africa.The foot switch was visually inspected upon receipt and was found to be in perfect working order.The device was functionally tested and passed all the tests during evolution.No other anomalies were identified.Therefore, the investigation is determined to be unconfirmed for the reported unintended movement issue.The root cause for reported unintended movement issue cannot be determined as the problem could not be reproduced.Labeling review: the labelling/packaging review was not required as the reported event is not associated with a labelling, packaging, or use related issue.D2b: (gei; knw), d4(unique identifier (udi) #), g2, g3, h6 (result, conclusion).Section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that during an upright vertical stereotactic procedure, the encor driver would not stop sampling after the user removed their foot off the encor foot pedal.The encor enspire needed to be switched off to stop the driver from sampling.After the enspire was switched back on, error code 1005 (driver button or contact stuck during initialization) was displayed.The enspire was switched off before removing the needle from the patient.There was patient involvement with no injury.It was identified that the encor enspire, the encor foot pedal and the encor driver were used together as a system during this event.
|
| |
|