| Catalog Number |
8187501 |
| Medical Device Problem Code |
No Flow (2991)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There has been a previous report received where lack of water flow has caused an overheating insert.Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
While using a cavitron jet plus g137 they allege that they have no water and the handpiece is getting hot, no injury resulted.
|
| |
|
Additional Manufacturer Narrative
|
|
10/28/25 evaluation tech: (b)(6).Emh hp; cable,conn/gun cable damaged.Damaged water flow control on the handpiece cable.Debris buildup in the water supply hose, wrong or missing water filter affecting water flow.Will replace damaged/worn components and recalibrate unit to factory specs upon estimate approval.No jm was sent in for evaluation.Hpc-08200.
|
| |
|
Search Alerts/Recalls
|