Catalog Number 366594
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when using one (1) bd microtainer® contact-activated lancet, the blade broke off in the customer's finger and was removed using tweezers.No medical invention was reported.No adverse impact was reported regarding the patient or user.
 
Additional Manufacturer Narrative
E1: initial reporter phone #: (b)(6).The manufacturing location for this product is htl.This site is an oem manufacturing site.Therefore, bd corporate headquarters in franklin lakes, nj has been listed in sections d.3.And g.1.And the franklin lakes fda registration number has been used for the manufacture report number.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
Investigation summary: bd had not received samples or photos for investigation.Therefore, retention samples from bd inventory were evaluated by visual examination and no issues were observed relating to breaks off during use as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode breaks off during use.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that when using one (1) bd microtainer® contact-activated lancet, the blade broke off in the customer's finger and was removed using tweezers.No medical invention was reported.No adverse impact was reported regarding the patient or user.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® CONTACT-ACTIVATED LANCET
Common Device Name
MANUAL SURGICAL INSTRUMENT FOR GENERAL USE
Manufacturer (Section D)
BECTON DICKINSON & CO (FRANKLIN LAKES)
1 becton drive
franklin lakes NJ 07417
Manufacturer (Section G)
BECTON DICKINSON & CO (FRANKLIN LAKES)
1 becton drive
franklin lakes NJ 07417
Manufacturer Contact
anand singh
1 becton drive
franklin lakes, NJ 07417
2018475186
MDR Report Key23310660
Report Number2243072-2025-01258
Device Sequence Number5801660
Product Code FMK
UDI-Device Identifier30382903665946
UDI-Public(01)30382903665946(17)290415(10)E7C78C9
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K223243
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number366594
Device Lot NumberE7C78C9
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2025
Supplement Date Received by Manufacturer11/21/2025
Initial Report FDA Received Date10/16/2025
Supplement Report FDA Received Date12/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/27/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-