• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 EXTRACTOR HANDLE; ORTHOPEDIC HANDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 EXTRACTOR HANDLE; ORTHOPEDIC HANDLE Back to Search Results
Catalog Number 17201A
Medical Device Problem Codes Break (1069); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary- according to the information provided: "it was reported that threaded t handle broke off with threads".The product was returned to depuy synthes for evaluation.Visual inspection of the returned device found the tip of the device fractured and cross-threaded.Broken fragments were not returned for evaluation.Device had signs of usage and no other anomaly was identified on its surface.Device was sent to the materials team for further investigation.Digital microscope inspection of the extractor handle threaded connection revealed thread damage.Visual inspection of the extractor handle threads also showed some threads¿ flanks had roughened as well as highly worn regions on the extractor handle shaft.It is evident that high stresses were applied during extraction loading of the extraction handle construct (load applied with either a mallet or slide hammer).Additionally, the combination of thread damage and worn extractor handle shaft regions indicate movement between the threaded interface during the loading application, which is likely to have occurred if the extractor handle and the threaded t-handle threaded connection was not fully engaged.In summary, the root cause of failure can be attributed to a combination of potential factors: a high user input load and partially seated threaded connection interface, not all of which are fully understood at this time, which resulted in loading conditions that exceeded the local shear strength of the material.This condition contributed to the thread deformation and fracture events.However, without concrete evidence or further information, it is not possible to determine a root cause.A dimensional inspection was unable to be performed due to the post-manufacturing damage.A functional test was unable to be performed due to the post-manufacturing damage.The overall complaint was confirmed as the observed condition of the extractor handle would have contributed to a device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot- the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.H11 additional narrative: added: h6.Medical device problem code corrected: h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that threaded t handle broke off with threads.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR HANDLE
Common Device Name
ORTHOPEDIC HANDLE
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23310701
Report Number1818910-2025-17663
Device Sequence Number5827259
Product Code LXH
UDI-Device Identifier00197169172010
UDI-Public(01)00197169172010(10)PG352567
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number17201A
Device Lot NumberPG352567
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/25/2025
Supplement Date Received by Manufacturer10/23/2025
12/22/2025
Initial Report FDA Received Date10/16/2025
Supplement Report FDA Received Date10/24/2025
12/24/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/21/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age59 YR
Patient SexMale
-
-