SMITH & NEPHEW, INC. ANTHOLOGY SO POROUS SZ 12; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE Back to Search Results
Catalog Number 71356012
Medical Device Problem Codes Delivered as Unsterile Product (1421); Tear, Rip or Hole in Device Packaging (2385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, one (1) anthology so porous sz 12 broke through packaging.As this was noticed upon inspection, there was no patient involvement.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The photographs were reviewed and revealed that the device broke through the packaging.However, a review made by the quality engineering team revealed that there are several possible failure modes that could have produced this output, and it is not possible to confidently identify the root cause without a returned product.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.According to the packaging sequence, the stem should be placed in a polyethylene bag with the distal end inserted first.This bag must then be placed inside the inner tray and sealed.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.Factors that could contribute to the reported event include damage during shipping or mishandling.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY SO POROUS SZ 12
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/CERAMIC/METAL, CEMENTED OR UNCE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key23310758
Report Number1020279-2025-01719
Device Sequence Number12009197
Product Code MRA
UDI-Device Identifier03596010567826
UDI-Public3596010567826
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K211176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number71356012
Device Lot Number18BM10777
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/18/2025
Supplement Date Received by Manufacturer11/28/2025
Initial Report FDA Received Date10/16/2025
Supplement Report FDA Received Date12/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/15/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-