RESPIRONICS, INC. DREAMSTATION 2; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number BLX520H15C
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/07/2025
Type of Reportable Event Malfunction
Event or Problem Description
A dreamstation 2 was returned to the third-party service center for evaluation.During the evaluation, it was noted that the device pca was burnt.In conclusion, the device pca needs to be replaced to address the issue.The root cause is confirmed at this time.Additionally, during the service of the device the inlet and outlet seal is missing and the pollen filter needs to be replaced for preventative maintenance unrelated to the reportable malfunction.The device has been forwarded to the manufacturer for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION 2
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23311895
Report Number2518422-2025-110957
Device Sequence Number12212866
Product Code BZD
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K200480
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation 505
Type of Report Initial,Followup
Report Date (Section B) 07/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberBLX520H15C
Device Catalogue NumberBLX520H15C
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/07/2025
Supplement Date Received by Manufacturer07/14/2026
Initial Report FDA Received Date10/16/2025
Supplement Report FDA Received Date07/14/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/20/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-