• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. DRIVER SHAFT, T-15, SHORT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. DRIVER SHAFT, T-15, SHORT; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number DRIVER SHAFT, T-15, SHORT
Medical Device Problem Codes Break (1069); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/06/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On 10/06/2025, a facility representative reported via (b)(4) that an ar-9545-t15-01 t-15 short drive shaft twisted and broke during a case.No patient was affected.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user-applied excessive force during the tightening process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRIVER SHAFT, T-15, SHORT
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key23312114
Report Number1220246-2025-04647
Device Sequence Number4967145
Product Code LXH
UDI-Device Identifier00888867225183
UDI-Public(01)00888867225183(11)200728(10)8002030
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 02/02/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDRIVER SHAFT, T-15, SHORT
Device Catalogue NumberAR-9545-T15-01
Device Lot Number8002030
Was Device Available for Evaluation? No
Distributor Facility Aware Date10/06/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2025
Supplement Date Received by Manufacturer01/29/2026
Initial Report FDA Received Date10/16/2025
Supplement Report FDA Received Date02/02/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/28/2020
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-