• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANALOG DEVICES SENSINEL CARDIOPULMONARY MANAGEMENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANALOG DEVICES SENSINEL CARDIOPULMONARY MANAGEMENT SYSTEM Back to Search Results
Model Number ADCP1100
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2025
Type of Reportable Event Malfunction
Event or Problem Description
During routine data monitoring on (b)(6) 2025, the team noticed an erratic pattern in some measurement signals.On (b)(6) 2025, the team confirmed inaccurate readings in 4 out of 18 diastolic heart sound strength measurements (22.22%).Two abnormal signal patterns were observed: (1) a shift in the acoustic signal away from the expected baseline and (2) abrupt simultaneous jumps/falls in both acoustic and impedance signals.These anomalies did not align with common user errors or poor device contact, suggesting a potential hardware malfunction.
 
Additional Manufacturer Narrative
During routine data monitoring on (b)(6) 2025, the team noticed an erratic pattern in some measurement signals.On (b)(6) 2025, the team confirmed inaccurate readings in 4 out of 18 diastolic heart sound strength measurements (22.22%).Two abnormal signal patterns were observed: (1) a shift in the acoustic signal away from the expected baseline and (2) abrupt simultaneous jumps/falls in both acoustic and impedance signals.These anomalies did not align with common user errors or poor device contact, suggesting a potential hardware malfunction.On (b)(6) 2025, analog notified the patient and the healthcare provider of the identified issue and advised the patient to discontinue using the device.In coordination with the clinical site, a return material authorization was issued, and the patient was given a replacement device on (b)(6) 2025, following established procedures.There were no adverse events or no need for medical intervention related to this malfunction.When the original unit was returned to analog on (b)(6) 2025, it underwent standard return processing.X-ray analysis confirmed a break in the wire in the acoustic signal path.This explained the shifting acoustic signal baseline and intermittent simultaneous jumps/falls caused by the wire making and breaking contact with patient movement.The root cause was confirmed to be a hardware failure due to ta broken connection in the acoustic signal path.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSINEL CARDIOPULMONARY MANAGEMENT SYSTEM
Common Device Name
CARDIOPULMONARY MANAGEMENT SYSTEM
Manufacturer (Section D)
ANALOG DEVICES
1 analog way
wilmington MA 01887
Manufacturer (Section G)
NEXT PHASE MEDICAL DEVICES
110 forbes blvd
mansfield MA 02048
Manufacturer Contact
sam rajkumar
1 analog way
wilmington, MA 01887
MDR Report Key23314373
Report Number3000130829-2025-00007
Device Sequence Number14694308
Product Code MWI
UDI-Device Identifier00860006382322
UDI-Public(01)00860006382322(17)270114
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K243216
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Biomedical Engineer
Remedial Action Inspection,Notification,Replace
Type of Report Initial
Report Date (Section B) 07/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberADCP1100
Device Lot Number30725820
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/18/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2025
Initial Report FDA Received Date10/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/14/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age66 YR
Patient SexUnknown
-
-