BOSTON SCIENTIFIC CORPORATION DYNAGEN? CRT-D; CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) Back to Search Results
Model Number G150
Medical Device Problem Codes Pacing Problem (1439); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code Shock from Patient Lead(s) (3162)
Date of Event 08/05/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the health care professional (hcp) called technical services (ts) for a consult review of this cardiac resynchronization therapy defibrillator (crt-d) stored ventricular episodes to confirm device operation.It was noted that the patient was in an atrial fibrillation (af) with rapid ventricular response (rvr) arrhythmia.Upon review, the ventricular episodes had recorded delivery of anti-tachycardia pacing (atp) and shock therapies.It was determined that the device had inappropriately delivered the therapy.The therapy was exhausted.Ts discussed programming options with the hcp.No adverse patient effects were reported.At this time, the crt-d remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYNAGEN? CRT-D
Common Device Name
CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key23316977
Report Number2124215-2025-73985
Device Sequence Number14694424
Product Code NIK
UDI-Device Identifier00802526534614
UDI-Public(01)00802526534614(17)200221
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
P010012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/21/2020
Device Model NumberG150
Device Catalogue NumberG150
Device Lot Number489832
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2025
Initial Report FDA Received Date10/16/2025
Date Device Manufactured03/06/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age73 YR
Patient SexMale
-
-