| Catalog Number |
625212S |
| Medical Device Problem Codes |
Positioning Failure (1158); Device Difficult to Setup or Prepare (1487)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information has been requested and if received, will be provided in the supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "the screw did not lock properly at the second hole from the distal end of the plate on the ulnar side.".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.If the device is returned or if any additional information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "the screw did not lock properly at the second hole from the distal end of the plate on the ulnar side.".
|
| |
|