• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH GAMMA3 - U-BLADE SET, TI D10.5X95MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH GAMMA3 - U-BLADE SET, TI D10.5X95MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number 30660095S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Perforation (2001)
Date of Event 09/18/2025
Type of Reportable Event Serious Injury
Event or Problem Description
As reported: "over-telescoping of the lag screw due to non-union.The screw does not have a cutout.The lag screw protruded to lateral due to collapse of the femoral neck.The patient required a revision surgery.".
 
Additional Manufacturer Narrative
Additional information has been requested and if received, will be provided in the supplemental report.
 
Event or Problem Description
As reported: "over-telescoping of the lag screw due to non-union.The screw does not have a cutout.The lag screw protruded to lateral due to collapse of the femoral neck.The patient required a revision surgery.".
 
Additional Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If more information is provided, the case will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GAMMA3 - U-BLADE SET, TI D10.5X95MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
dannielle stap
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23318451
Report Number0008031020-2025-01943
Device Sequence Number4989278
Product Code HSB
UDI-Device Identifier04546540595997
UDI-Public(01)04546540595997(11)211112(17)261031(10)K043F59
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K043431
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Catalogue Number30660095S
Device Lot NumberK043F59
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/18/2025
Supplement Date Received by Manufacturer11/20/2025
Initial Report FDA Received Date10/16/2025
Supplement Report FDA Received Date12/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-