|
An event regarding a product mix involving an exeter stem was reported.The event was confirmed via evaluation of the box, label and part.Method & results -product evaluation and results: a visual inspection was conducted on the device, label and packaging.The catalog and lot number on the box and label (0580-1-442, lot a00976) did not correspond to the catalog and lot number laser marked on the stem (0580-1-352, lot g8754849).Material analysis, functional and dimensional inspections were not be performed as they were not relevant to the event.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the product history records indicate the reported device was manufactured with no reported relevant discrepancies.-complaint history review: there have been 11 other similar events for the reported lot, all related to the same issue.Conclusions: it was reported 'in addition to the incorrect stem that was cemented into the patient, we have found more incorrect stems (37.5/0) in the (b)(4) boxes lot number: a00976.We have found another x (b)(4) so far ((b)(4) total) all have been taken off the shelf.' a visual inspection was conducted on the device, label and packaging.The catalog and lot number on the box and label (0580-1-442, lot a00976) did not correspond to the catalog and lot number laser marked on the stem (0580-1-352, lot g8754849).This event is deemed to be a manufacturing issue.Manufacturing nc was issued on 15 aug 2025 for this event.
|