MAUDE Adverse Event Report: EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE MITRAL PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE MITRAL PERICARDIAL BIOPROSTHESIS; REPLACEMENT HEART-VALVE Back to Search Results
Model Number 7300TFX27
Medical Device Problem Code Calcified (1077)
Health Effect - Clinical Code Fatigue (1849)
Date of Event 09/23/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
H11: additional manufacturer narrative: this report serves as both the initial and final medical device report (mdr), incorporating the findings from the investigation.The device was not returned to edwards for evaluation as it remains implanted.The instructions for use (ifu) have been reviewed, and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.The reported type of event is included in the ifu.The device history record review was completed, and this device passed all manufacturing and sterilization inspections.No nonconformances related to the complaint event were identified.Svd (structural valve degeneration) is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv (bioprosthetic heart valves) implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Despite the advancements in the heart valve industry, svd commonly begins 7- 8 years after implantation, with a considerably increased rate in patients with pertinent comorbidities and/or an implant duration of greater than 10 years.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including an implant duration of 7 or more years.Since no edwards defect was identified, corrective or preventative actions are not required.
 
Event or Problem Description
Edwards received notification from taiwan, that this 67-year-old patient with a valve model 7300tfx27 implanted in the mitral position underwent a valve-in-valve intervention after an implant duration of at least seven (7) years and eight (8) months due to calcification, which led into mitral stenosis.The patient presented with fatigue before the procedure.A transcatheter valve was implanted in replacement, and the patient was noted as to be safe and stable.The implant date is only known as "2017".Therefore, (b)(6) 2017 is being used to calculate the minimum implant duration.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE MITRAL PERICARDIAL BIOPROSTHESIS
Common Device Name
REPLACEMENT HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 office m2013
irvine, CA 92614
9492506615
MDR Report Key23322035
Report Number2015691-2025-08561
Device Sequence Number21309761
Product Code DYE
UDI-Device Identifier00690103186260
UDI-Public(01)00690103186260(17)201010
Combination Product (Y/N)N
Initial Reporter CountryTW
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Expiration Date10/10/2020
Device Model Number7300TFX27
Device Lot Number5176698
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2025
Initial Report FDA Received Date10/17/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/11/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age67 YR
Patient SexMale
-
-