• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC. INFUSOMAT®; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 362035
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: details of complaint (reported issue): "b-braun pumps alarming for air in line, lines reprimed, no air in line per 2 nurse visually checking, lines changed, still alarming, pumps changed.For a fentanyl infusion and ns push infusion." impact to patient: delay in treatment, interruptions to clinical care due to time required to resolve alarms.Action(s) taken changed out pump (isolate and send to bioengineering department), followed b braun tips to resolve alarms (air-in-line and occlusion), repeated priming to clear air, tubing replaced (triple bag and send to material management), other.Medication / fluid ns iv rate 15cc/hr.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation, the root cause of the reported incident was unable to be determined.The reported defect was unable to be confirmed.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Additionally, the retained units were evaluated and passed the internal testing.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFUSOMAT®
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th ave
bethlehem PA 18018
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th ave
bethlehem PA 18018
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key23322463
Report Number2521402-2025-01182
Device Sequence Number9949044
Product Code FPA
UDI-Device Identifier04046964699047
UDI-Public(01)04046964699047(11)240918(17)290831(10)0061962760
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number362035
Device Lot Number0061962760
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2025
Supplement Date Received by Manufacturer10/02/2025
Initial Report FDA Received Date10/17/2025
Supplement Report FDA Received Date11/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-