|
Additional information: d9, g3, h3, h6.The reported event was confirmed as one unpackaged ar-13997mff, fastpass scorpion-mf with flushport, batch number 15051240, was received for investigation and upon visual inspection, it was observed that the tip of the upper moving jaw is broken off (see evaluation picture 3).Functional testing was not performed due to the damage of the device.No fragments were returned for evaluation.The most likely cause for the reported failure can be attributed to applying excessive force while grasping and manipulating tissue / suture, or when applying excessive leveraging forces.Further the device shows signs of metal surface oxidation at the tip.
|