• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP FORCEPS; FORCEPS, OPHTHALMIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER REVOLUTION DSP FORCEPS; FORCEPS, OPHTHALMIC Back to Search Results
Catalog Number 705.44P
Medical Device Problem Code Activation Problem (4042)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported that, during vitrectomy surgery an ophthalmic forceps did not work correctly.The surgery was completed on the same day by opening new forceps without any patient harm.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information was provided in sections d.9, h.3, h.6 and h.11.The returned instrument was received in its outer blister, covered by the tyvek foil lid, shows the lot information.The inner blister which protects the instrument during the shipment was not used.The product shows macroscopic sign of damage, namely that the forceps arms are significant bent.There are no surgery residues.The instrument was visually inspected with the aid of a photomicroscope with various magnifications.The visual inspection shows the bent arms in detail.The level of damage as well as the missing inner blister suggest that the deformation was caused during the shipping process from the complainant to the investigation site.However, the point in time when the described damage occurred can no longer be determined conclusively nor can the situation that led to it be reconstructed.The customer¿s complaint is confirmed as the returned device is damaged, but a root cause for the reported issue cannot be determined.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Non-conformance report provided by originator.A non-conformance-based review of the lot number was performed and did not reveal any potential contributing factors to the reported complaint it cannot be determined how or where the damage to the sample occurred.Therefore, the root cause for the customer's reported event could not be determined.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all manufacturer products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.The complaint data will continue to be monitored for evidence of adverse trending and further action will be taken, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER REVOLUTION DSP FORCEPS
Common Device Name
FORCEPS, OPHTHALMIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23323671
Report Number3003398873-2025-00391
Device Sequence Number13415008
Product Code HNR
UDI-Device Identifier07612717070622
UDI-Public(01)07612717070622(17)280331(11)250417(10)17D1F5
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number705.44P
Device Lot Number17D1F5
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/24/2025
Supplement Date Received by Manufacturer10/29/2025
Initial Report FDA Received Date10/17/2025
Supplement Report FDA Received Date11/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/17/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-