Catalog Number 05.001.083
Medical Device Problem Codes Break (1069); Leak/Splash (1354); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the double air hose, 3m, for air pen drive device had a leakage noise when connecting the hose to nitrogen.Motor device overheated.It was reported that it seemed to be by a rupture or failure of the internal hose to the coupling on the end of it.It was not reported if the device was used in surgery.It was not reported if there were any delays in a surgical procedure, or if a spare device was available for use.There were no reports of injuries, medical intervention or prolonged hospitalization." all available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the device has not been previously serviced.Device history record shows the serial number of 05.001.083 was processed through the normal manufacturing and inspection operations with no rework or nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacturing process.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: h4: updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOUBLE AIR HOSE, 3M, FOR AIR PEN DRIVE
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 
SZ  
3035526892
MDR Report Key23325283
Report Number8030965-2025-10480
Device Sequence Number9909905
Product Code HBB
UDI-Device Identifier07611819831520
UDI-Public(01)07611819831520
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K093361
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number05.001.083
Device Lot Number34535
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/10/2025
Supplement Date Received by Manufacturer12/22/2025
12/24/2025
Initial Report FDA Received Date10/17/2025
Supplement Report FDA Received Date12/23/2025
12/24/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/10/2021
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-