ALERE SAN DIEGO, INC. MEDLINE HCG COMBO+ PREGNANCY TEST CASSETTE (SERUM/URINE); VISUAL, PREGNANCY HCG, PRESCRIPTION USE
|
Back to Search Results |
|
| Model Number |
FHC-A202 |
| Medical Device Problem Code |
False Positive Result (1227)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3 - date of event: was not provided, therefore 1sep2025 was selected.Investigation conclusion: retained devices from the reported lot number were tested with hcg-negative clinical urine samples.The results were read at 3 and 4 minutes and all devices yielded valid negative results with strong control lines.No migration issues were observed during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review did not uncover any relevant non-conformances and found that the lot met quality control specifications.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the package insert: very low levels of hcg (less than 50 miu/ml) are present in serum and urine specimen shortly after implantation however, because a significant number of first trimester pregnancies terminate for natural reasons, a test result that is weakly positive should be confirmed by retesting with a first morning serum or urine specimen collected 48 hours later.A number of conditions other than pregnancy, including trophoblastic disease and certain non-trophoblastic neoplasms including testicular tumors, prostate cancer, breast cancer, and lung cancer, cause elevated levels of hcg.Therefore, the presence of hcg in a serum or urine specimen should not be used to diagnose pregnancy unless these conditions have been ruled out.As with any assay employing mouse antibodies, the possibility exists for interference by human anti-mouse antibodies (hama) in the specimen.Specimens from patients who have received preparations of monoclonal antibodies for diagnosis or therapy may contain hama.Such specimens may cause false positive or false negative results.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
|
| |
|
Event or Problem Description
|
|
The distributor reported that a customer received 5 false positive hcg results while using medline hcg combo cassettes on 4 patients.Although requested, it is unclear which patient experienced 2 false positive hcg results.No adverse health outcomes were reported for all patients.The customer reported some of the patients had no sexual intercourse, however no further information was provided.See mdrs 2027969-2025-00211, 2027969-2025-00213, and 2027969-2025-00214 for patients 1,3, and 4 respectively.
|
| |
|
Search Alerts/Recalls
|
|
|