|
Ge healthcare (gehc) reported a field modification for reduced concentration delivery of anesthetic agent per 21 cfr 806 on 27-nov-2024.The fda recall number is (recall no.Z-0814-2025).Customers were sent a letter explaining the issue and instructions for inspecting for correct output concentration.Gehc will replace all affected units.Block a: no report of patient involvement.Legal manufacturer: hcs madison: 3030 ohmeda dr, usa madison, wi 53718.
|
|
As a result of an inspection that was completed as part of the correction initiated by ge healthcare (gehc) on 02-apr-2025, (recall numbers z-1624-2025, z-1625-2025, z-1626-2025, z-1627-2025, z-1628-2025, z-1629-2025, z-1630-2025, z-1631-2025, z-1632-2025, z-1633-2025, z-1634-2025, z-1635-2025, z-1636-2025, z-1637-2025, z-1638-2025) this unit was identified as having an issue that prevents effective mechanical ventilation in volume control ventilation (vcv) mode.There was no patient involvement.
|