• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. PWRD ECH FLEX 45MM 34CM SHAFT; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. PWRD ECH FLEX 45MM 34CM SHAFT; STAPLE, IMPLANTABLE Back to Search Results
Catalog Number PSE45A
Medical Device Problem Code Ejection Problem (4009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Date sent: 10/17/2025.D4: batch #660d72.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one pse45a device was returned with no visual non-conformances and without a reload present.In addition, a reload pan was found lodged inside the channel.The pan was removed from the channel and the device was tested for functionality with a test reload.The device fired, cut, and formed all the staples as intended.The staple line and cut line were complete and the staples met the staple release criteria.It is also possible that handling by the customer could dislodge the pan.Dislodgement as a result of the removal of the reload from the device is the most likely scenario.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot/batch number, a98e0y, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 660d72, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during a laparoscopic lobectomy, it was observed that the cartridge could not be installed.Changed to a new one to complete surgery.There was no patient consequence reported.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PWRD ECH FLEX 45MM 34CM SHAFT
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 00969
*   00969
3035526892
MDR Report Key23327910
Report Number3005075853-2025-08099
Device Sequence Number9910451
Product Code GDW
UDI-Device Identifier10705036014317
UDI-Public(01)10705036014317(10)A98E0Y(17)280630
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110385
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/17/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPSE45A
Device Lot NumberA98E0Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2025
Initial Report FDA Received Date10/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-