It was reported that during service and repair pre-testing, it was discovered that the neuro-tip of the craniotome device was bent/damaged.It was further determined that the device met the visual assessment criteria.This event did not occur during surgery.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2025.All available information has been disclosed.
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H11: additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the craniotome neuro-tip being bent/damaged, identified during service and evaluation was confirmed.The assignable root cause was determined to be due to component failure from wear.
|