SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD ALARIS SECONDARY SET; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 11610341
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd secondary set was leaking.The following information was received by the initial reporter with the following verbatim.It was reported by customer that they noticed there were 5-6 drops on the floor and there was leakage around the texium that was connected to the primary tubing.
 
Additional Manufacturer Narrative
No product or photo was returned by the customer.The customer complaint of leakage could not be verified due to the product not being returned for failure investigation.A device history record review for material# 11610341 and lot# 24125087 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on 04dec2024.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.Due to no sample being received, an investigation could not be performed, and a root cause could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SECONDARY SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key23333973
Report Number9616066-2025-03180
Device Sequence Number10536601
Product Code FPA
UDI-Device Identifier10885403221842
UDI-Public(01)10885403221842(17)271204(10)24125087
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K022209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number11610341
Device Lot Number24125087
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2025
Supplement Date Received by Manufacturer10/22/2025
Initial Report FDA Received Date10/19/2025
Supplement Report FDA Received Date10/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/04/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-