• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® DRYSEAL FLEX INTRODUCER SHEATH; INTRODUCER, CATHETER Back to Search Results
Catalog Number DSF2433
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 10/01/2025
Type of Reportable Event Serious Injury
Event or Problem Description
The following information was reported to gore: on (b)(6) 2025, the patient underwent endovascular treatment of a thoracic aortic aneurysm, zone 2, using gore® tag® thoracic branch endoprostheses.A 24 fr gore® dryseal flex introducer sheath was inserted via the right common femoral artery.After placement of the stent graft and removal of the sheath, an access injury was identified a few centimeters proximal to the puncture site in the right common femoral artery.As a treatment, vascular graft replacement was performed.The diameter of the right access vessel was approximately 7.8 mm to 9.2 mm.
 
Additional Manufacturer Narrative
H6: a review of the manufacturing records for the device could not be performed as a valid lot number was not provided.H6: d1102, ifu statements the gore® dryseal flex introducer sheath instructions for use (ifu) was reviewed and the following ifu statements were identified in relation to the reported event.Warnings: adequate vessel access is required to introduce the sheath into the vasculature.Careful evaluation of vessel size, anatomy, tortuosity, and disease state (including calcification, plaque, and thrombus) is required to ensure successful sheath introduction and subsequent withdrawal.If vessel is not adequate for access, major bleeding, vessel damage, or serious injury to the patient, including death, may result.¿ if vessel size is smaller than the nominal body od (table 1), major bleeding, vessel damage, or serious injury to the patient, including death, may result.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® DRYSEAL FLEX INTRODUCER SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
rie kaba
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key23334268
Report Number3007284313-2025-04320
Device Sequence Number18784825
Product Code DYB
UDI-Device Identifier00733132630950
UDI-Public(01)00733132630950(17)260502
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K160254
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/02/2026
Device Catalogue NumberDSF2433
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2025
Initial Report FDA Received Date10/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexMale
-
-