An event regarding instability/wear involving a gmrs tibial component was reported.The event was not confirmed.Method & results: -product evaluation and results: the reported devices were not returned; however photographs were provided for review.The photographs show the femoral, tibial, and rotating components after explantation.The poly bearing components appear fractured and/or worn.There is no damage observed on this gmrs tibial component (6495-3-101) in the images provided.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the product history records indicate the reported device was manufactured with no reported relevant discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to instability and wear of various components.The reported devices were not returned; however photographs were provided for review.The photographs show the femoral, tibial, and rotating components after explantation.The poly bearing components appear fractured and/or worn.There is no damage observed on this gmrs tibial component (6495-3-101) in the images provided.Further information such as return of the devices, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|