STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON P/A CR BEADED #7L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Catalog Number |
5517F701 |
| Medical Device Problem Codes |
Unstable (1667); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Joint Dislocation (2374); Joint Laxity (4526)
|
| Date of Event |
09/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patella dislocated post total knee replacement which was performed on (b)(6) 2025.No implants were revised.Surgeon was able to do a lateral release to gain stability without changing any implants.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding instability involving a triathlon femoral component was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remains implanted.Clinician review: no medical records were provided for review with a clinical consultant.Product history review: review of the device history records indicate the devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
No new information.
|
| |
|