|
(b)(4).The actual device was not returned; however, the customer provided one photo for analysis.The photo shows the arterial luer hub with a crack attached to a syringe.A complete visual inspection could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.The instructions for use (ifu) provided with this kit instructs the user, "alcohol , alcohol-based solutions (e.G.Hibiclens, chloraprep) , iodine-based solutions (povidone-iodine) peg-based ointments (e.G., bactroban), hydrogen peroxide, or exsept plus are accepted for use with this catheter at the exit site (including the juncture hub and catheter body).Ensure that solution is completely dry before applying an occlusive dressing.Warning: avoid excessive or prolonged use of alcohol-based solutions and ointments to clean catheter or site care." the ifu also instructs the user, "examine catheter and extension lines before and after each treatment for any signs of damage." the customer report of a cracked luer hub was able to be confirmed by visual inspection of the customer supplied photo.Full complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Additional information was received.There was no air present in the catheter during, aspiration, flushing, or during dialysis treatment.The catheter was not leaking when flushed.The patient had the catheter for about two months.The operator did not overtighten the bloodlines, syringes, or caps.Sustaining engineering was contacted as part of this complaint investigation and it was determined that the observed damage is consistent with damage due to alcohol exposure.A non-conformance has been initiated to further investigate this issue.Based on the customer report and the supplied photo, the probable cause is likely design related.Teleflex will continue to monitor and trend for reports of this nature.
|