Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D3: email adress.D4: udi number: corrected.D4: expiration date: updated to unknown.G1: contact office - first/given name, last name and email address.G1: contact office - manufacturer site - email address.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) tascd30, (tsv angled screw channel driver 30mm) for evaluation.Visual evaluation was performed; the driver has signs of use.The driver¿s tip is fractured.The fractured piece was not returned.Dhr review was completed for the subject lot number 1282451.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 1282451 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental: functional: fracture¿ based on the investigation and risk management file review, the most likely root cause determined from the investigation was customer error, applying too much torque.Therefore, based on the available information, a device malfunction did occur.The driver¿s tip was fractured.The reported event was confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|