| Model Number |
8886803512 |
| Medical Device Problem Code |
Entrapment of Device (1212)
|
| Health Effect - Clinical Codes |
Unspecified Tissue Injury (4559); Insufficient Information (4580)
|
| Date of Event |
08/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 - concomitant product: 8886803512, 8886803512 appose ulc 35 reg skin stap, (lot# j4l2795ly).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during the stapling of the laceration to the back of the head, the skin stapler malfunctioned and got stuck to the patient while attached to the staple.After much adjustment and when lidocaine was used to numb the patient, the surgeons were able to get the stapler removed from the head.The damaged skin staples were removed.The first stapler misfired at staple number four.A second stapler also had the same malfunction on staple number seven.Another skin stapler was used to complete the case.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the skin stapler malfunctioned and got stuck to the patient while attached to the staple.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|