| Model Number |
SR-0940-CS |
| Medical Device Problem Codes |
Nonstandard Device (1420); Defective Device (2588)
|
| Health Effect - Clinical Codes |
Ischemia Stroke (4418); Thrombosis/Thrombus (4440)
|
| Date of Event |
09/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the stent was compressed, and the patient experienced a stroke.An enroute transcarotid stent was selected for use in a left sided transcarotid artery revascularization (tcar) procedure.The target vessel was significantly calcified.During the tcar procedure, there was inadequate imaging and post-dilation as a result of patient agitation.Approximately 45 minutes post procedure, the patient experienced an ischemic stroke and imaging identified calcium compressing the stent.Neurology interventional radiologists performed a thrombectomy and successfully post-dilated the stent.No further information was provided to boston scientific.
|
| |
|
Event or Problem Description
|
|
It was reported that the stent was compressed, and the patient experienced a stroke.An enroute transcarotid stent was selected for use in a left sided transcarotid artery revascularization (tcar) procedure.The target vessel was significantly calcified.During the tcar procedure, there was inadequate imaging and post-dilation as a result of patient agitation.Approximately 45 minutes post procedure, the patient experienced an ischemic stroke and imaging identified calcium compressing the stent.Neurology interventional radiologists performed a thrombectomy and successfully post-dilated the stent.No further information was provided to boston scientific.
|
| |
|