| Model Number |
23AGFN-756 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a 23mm sjm regent heart valve w/ flex cuff was chosen for a procedure.After the implantation of valve, the leaflet was noted sticking in the tissue and it was opening but not properly.The physician explanted the valve immediately and implanted a non-abbott valve while the patient was still on bypass.There was no delay in the procedure.The patient was reported stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of leaflet sticking in tissue was reported.The valve was sent for functional testing at the engineering test lab.During this test, which ensures proper leaflet coaptation and hemodynamic performance, the valve was placed in a pulse duplicator filled with saline.The conditions included a beat rate of 70 ± 1 bpm, a flow rate of 5.0 ± 0.2 lpm, and a mean aortic pressure of 100 +10/-0 mmhg.Under these conditions, the leaflets appeared to coapt such that the effective orifice area value was 2.35 and the regurgitant fraction value was 9.6%.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, including functional testing to ensure proper leaflet coaptation and hemodynamic performance, and the product met all specifications.The cause of reported event could not be determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|