Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the instrument was fully applied.Further inspection noted that the green bar was not visible in the indicator window and the safety feature was not engaged.The staple guide was observed to be attached to the instrument with no damage to the staple guide.After actuating the handle, the pusher fingers appeared to be intact and inspection under the microscope displayed no damage to the knife blade.{anvil} the anvil of the instrument was observed to be not tilted and separated from the instrument.Further inspection of the anvil cutting ring showed no traces of knife impression and the ring was received intact.The wing nut and safety functioned properly upon rotating the twist knob.The green bar was visible within the indicator window when the twist knob was fully closed.The instrument was applied to the appropriate test media producing acceptable results.Pusher-fingers and knife advance when the handle was squeezed and the knife cut the media cleanly and completely.The device also produced all audible, tactile and visual indicators during the functional testing.The instrument body label was removed for evaluation of the eccentric washer assembly.Evaluation of the eccentric washer noted that it was secured.It was reported that the circular stapler did not fire during use.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The evaluation detected an unreported condition: there was no traces of knife impression.The product analysis noted evidence that the device was not used as intended.This issue can occur if the instrument handle compression is not completed or if the instrument is applied against an excessive amount of tissue during the clinical application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: failure to squeeze the handle fully during firing may result in unacceptable staple formation and/or an incomplete knife cut.This may result in leakage.Ensure the handle is fully squeezed when the metal underside of the handle contacts the stapler body to the fullest extent.Also, when firing the stapler, make sure to fully squeeze the instrument handle until the metal underside of the handle contacts the stapler body to the fullest extent.Partial or incomplete handle squeezes may result in unacceptable staple formation and/or incomplete knife cut.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|