| Catalog Number |
45-400 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
05/01/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side ¿rupture¿.Device was explanted and replaced.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side ¿rupture¿.Device was explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ rupture: observed an opening assessed as fold flaw opening.As per the investigation procedure, creases, wear abrasion and stress marks were observed and none of the observations are found to be potentially related to the manufacturing process, no further actions are required.
|
| |
|