| Catalog Number |
468-350 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/15/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "deflation and exchange from textured to smooth devices due to the patient's concern with the product".This record is for the left side.Device was explanted.Device will be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.A review of the device history record has been completed.No deviations or non-conformances noted.Reason for reoperation: deflation.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: deflation: observed two opening through microscopic inspection assessed as surgical damage.No further actions are required as it is not related to the manufacturing process.Anxiety-product/procedure: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (creases and wear abrasion) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: d.9., h.3., h.6.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation and exchange from textured to smooth devices due to the patient's concern with the product" and "any creases".This record is for the left side.Device was explanted.Device will be returned.
|
| |
|
Search Alerts/Recalls
|