| Catalog Number |
050-87215A |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Abdominal Pain (1685); Peritonitis (2252); Peritoneal effluent abnormal (4904)
|
| Date of Event |
10/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this patient with end stage renal disease (esrd) on continuous cyclic peritoneal dialysis [cc(pd)] utilizing a liberty select cycler and liberty cycler set for renal replacement therapy (rrt) was hospitalized for peritonitis.Follow-up with the patient¿s pd registered nurse (pdrn) confirmed the patient presented to the emergency room on (b)(6) 2025 with complaints of abdominal pain and cloudy peritoneal effluent fluid.A peritoneal effluent fluid culture and cell count (results unavailable) were collected, and the patient was admitted.The patient was diagnosed with peritonitis and was treated with (ip) vancomycin 2000 mg every 5 days, and ceftazidime 1500 mg daily for 14 days.The patient¿s peritoneal effluent culture result returned positive for kocuria rhizophila on (b)(6) 2025, and the patient¿s ceftazidime was discontinued.The patient continued to undergo while hospitalized without any reported issues and was discharged home in stable condition on (b)(6) 2025.The patient has recovered from the serious adverse events and continues to undergo ccpd therapy at home utilizing the same liberty select cycler without reported issues.The pdrn attributed causality to an unspecified touch contamination event based on the cultured organism.Per the pdrn, the serious adverse events were not caused by any fresenius product(s) and/or device(s) deficiency or malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.Clinical review: a temporal relationship exists between ccpd therapy utilizing a liberty cycler set and the serious adverse events of peritonitis, characterized by abdominal pain and cloudy peritoneal effluent fluid, which warranted hospitalization and antibiotic therapy.The pdrn attributed causality to an unspecified contamination event.Those individuals undergoing pd therapy (manual or cycler based) are at high risk for infections of the peritoneum.Kocuria microorganisms are ubiquitous and can be isolated from various sources, such as the skin, the flora of mammals, and soil.These microorganisms usually colonize the skin, mucosae, and oropharynx in humans.¿ according to the pdrn, the serious adverse events were related to a fresenius product(s) and/or device(s) defect or malfunction.Based on the information available, the patient¿s liberty cycler set can be disassociated from the serious adverse events.There is no allegation or objective evidence indicating a fresenius device(s) and/or product(s) caused or contributed to the serious adverse events.Furthermore, there is no report a fresenius device(s) and/or product(s) failed to meet the users¿ expectations and/or the manufacturers¿ specifications.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the patient for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius liberty cycler sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.There is no allegation of cassette malfunction based in the complaint description and information provided, the reported peritonitis was attributed to a breach in aseptic technique.There is no allegation of cassette malfunction.As there is no allegation related to product manufacture, the complaint is deemed unconfirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that this patient with end stage renal disease (esrd) on continuous cyclic peritoneal dialysis [cc(pd)] utilizing a liberty select cycler and liberty cycler set for renal replacement therapy (rrt) was hospitalized for peritonitis.Follow-up with the patient¿s pd registered nurse (pdrn) confirmed the patient presented to the emergency room on (b)(6) 2025 with complaints of abdominal pain and cloudy peritoneal effluent fluid.A peritoneal effluent fluid culture and cell count (results unavailable) were collected, and the patient was admitted.The patient was diagnosed with peritonitis and was treated with (ip) vancomycin 2000 mg every 5 days, and ceftazidime 1500 mg daily for 14 days.The patient¿s peritoneal effluent culture result returned positive for kocuria rhizophila on (b)(6) 2025, and the patient¿s ceftazidime was discontinued.The patient continued to undergo while hospitalized without any reported issues and was discharged home in stable condition on (b)(6) 2025.The patient has recovered from the serious adverse events and continues to undergo ccpd therapy at home utilizing the same liberty select cycler without reported issues.The pdrn attributed causality to an unspecified touch contamination event based on the cultured organism.Per the pdrn, the serious adverse events were not caused by any fresenius product(s) and/or device(s) deficiency or malfunction.
|
| |
|
Search Alerts/Recalls
|