• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC TSV BELLATEK ENCODE EMERGENCE HEALING ABUTMENT 5.7MM(D) 6.8MM(P) 5MM(H); DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC TSV BELLATEK ENCODE EMERGENCE HEALING ABUTMENT 5.7MM(D) 6.8MM(P) 5MM(H); DENTAL ABUTMENT Back to Search Results
Catalog Number TEEHA565
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided a2: age and date of birth unknown / not provided a3: patient sex unknown / not provided a4: weight unknown / not provided b3: date of event unknown / not provided e1: email address unknown / not provided it is unknown if the device will be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the healing abutment fractured off at the screw portion in the patients mouth.Requested screw removal tools in order to retrieve the broken portion from the implant.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111, 4114 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H10: narrative/data was updated.Zimvie did not receive the reported healing abutment for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was material selection of the product is not adequate to withstand occlusal forces.Therefore, based on the available information, device malfunction was not verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TSV BELLATEK ENCODE EMERGENCE HEALING ABUTMENT 5.7MM(D) 6.8MM(P) 5MM(H)
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos  gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23342473
Report Number0001038806-2025-02530
Device Sequence Number9451682
Product Code NHA
UDI-Device Identifier00844868045960
UDI-Public(01)00844868045960(17)290415(10)1280100
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K212730
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/25/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberTEEHA565
Device Lot Number1280100
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/20/2025
Supplement Report FDA Received Date11/25/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/16/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-