| Catalog Number |
367977 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There were multiple 510k numbers reported to be involved.The information is as follows: g.4.Pma / 510(k)#: k230855.Investigation summary: bd received 9 samples for investigation.The customer samples were subjected to a visual inspection for bowed tubes and 6 of 9 samples failed inspection.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint has been confirmed for the indicated failure mode bowed assembly.Bd was not able to identify a root cause for the indicated failure mode.Complaints received for this device and reported condition will continue to be tracked and trended.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® sst¿, an unspecified number of tubes were discovered to be bent post collection.No adverse impact was reported regarding the patient or user.
|
| |
|