• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9750TFX20A
Medical Device Problem Codes Gradient Increase (1270); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Edwards received notification from the implant patient registry that a patient with a 9750tfx 20mm sapien 3 ultra valve, implanted in the aortic position, underwent an explant procedure after an implant duration of 5 months and 22 days due to unknown reasons.A 23mm edwards surgical valve was implanted in replacement.
 
Additional Manufacturer Narrative
Per the instructions for use (ifu), non-emergent reoperation, emergency cardiac surgery, reoperation, device migration or malposition requiring intervention and device explant are listed as potential risks associated with the device and transcatheter valve replacement (tvr) procedure.The ifu cautions that long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.Accelerated deterioration of the valve due to calcific degeneration may occur in children, adolescents, or young adults and in patients with an altered calcium metabolism.Valve recipients should be maintained on anticoagulant/antiplatelet therapy, except when contraindicated, as determined by their physician.This device has not been tested for use without anticoagulation.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Valve reintervention may be due to valve dysfunction (e.G., severe, or clinically significant paravalvular leak, central leak, significant malposition, early/late endocarditis, thrombosis, structural valve deterioration or degeneration) and/or other reasons unrelated to the edwards devices (e.G., treatment of other patient comorbidities).During the manufacturing process, all sapien valves (all models) are 100% visually inspected for defects and 100% tested under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.In this case, despite multiple investigational attempts, it was not possible to obtain additional details regarding this event.Due to limited information, a definitive root cause was unable to be determined at this time.At the time of this report, the information available does not reasonably suggest there was a malfunction of the edwards device or that the use or miss-use of the device caused or contributed to the event.There is no allegation of deficiencies related to the identity, quality, durability, reliability, safety, labeling effectiveness, or performance of this edward device.Should additional information become available, the information will be reviewed, and the file updated accordingly.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
 
Event or Problem Description
Edwards received notification from the implant patient registry that a patient with a 9750tfx 20mm sapien 3 ultra valve, implanted in the aortic position, underwent an explant procedure after an implant duration of 5 months and 22 days due to unknown reasons.A 23mm edwards surgical valve was implanted in replacement.Per the medical records provided, the patient presented for tavr valve follow up and was found to have increased gradients.A ct of the heart and lhc indicated that the patient had severe obstructive disease in the lima to lad, svg to om1, and other heart vessels.The patient needed bypass surgery to correct the obstructions.
 
Additional Manufacturer Narrative
Supplemental report submitted to include additional information received through follow-up.Per the instructions for use (ifu), structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis), and device degeneration are known potential risks associated with bioprosthetic heart valves.The ifu cautions that accelerated deterioration of the valve may occur in patients with an altered calcium metabolism.Long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.It is normal to have a small gradient across a prosthetic valve after implant; however, elevated gradients may indicate obstructed flow across the valve.Residual gradients after implantation may result from patient factors such as hypertrophic cardiomyopathy (hcm), sub-valvular aortic stenosis, or inadequate leaflet coaptation.It may also be indicative of sub-optimal frame expansion or patient-prosthesis mismatch.Over time, gradients can develop or worsen due to leaflets being affected by the development of calcification, or the formation of pannus/host tissue, or thrombus.Low gradients may not be indicative of significant dysfunction.High gradients may be indicative of ongoing dysfunction and may require surgical or percutaneous intervention to restore normal flow.Structural valve deterioration (svd) may be manifested as stenosis with thickened leaflets.Svd may be mild and not require any intervention or it may be moderate to severe.It can cause the heart to work harder to eject blood from the ventricle.Depending on the severity it could be an indication for valve replacement or medical intervention.Tissue calcification is a common failure mode.The mechanisms for bioprosthetic heart valve tissue calcification are not yet fully understood.Many factors can contribute to the onset and propagation of calcification including patient-related (e.G., patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.During the manufacturing process, all sapien thv valves are 100% visually inspected for defects and 100% functionally tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.In this case, there was no allegation or indication a manufacturing non-conformance contributed to this product problem.Investigation results suggest that patient related factors caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 ULTRA TRANSCATHETER HEART VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key23343775
Report Number2015691-2025-08675
Device Sequence Number21197508
Product Code NPT
UDI-Device Identifier00690103201314
UDI-Public(01)00690103201314(17)270416
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
P140031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number9750TFX20A
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/25/2025
Supplement Date Received by Manufacturer10/29/2025
Initial Report FDA Received Date10/20/2025
Supplement Report FDA Received Date11/18/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/16/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexMale
-
-