| Catalog Number |
0167L22 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Initial reporter full facility name is (b)(6).The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.All available udi information is being provided.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon ruptured of the foley catheter and came out.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed - cause unknown.The product had caused the reported failure.Visual inspection noted sac burst along lumen without missing pieces.Introduced water into the inflation lumen and did not experience any obstructions to impede flow.Microscopic examination found there was present lubricant on the balloon area.However, the exact cause on how and when problem occurred could not be determined.The potential root cause for this failure could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/operator error).The labeling and instructions for use were found to be adequate.A review of the device history record did not show any problems or conditions that would have contributed to the reported issue.Therefore, no additional action is required at this time.Correction: h this report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon ruptured of the foley catheter and came out.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Initial reporter full facility name is osaka habikino medical center, osaka prefectural hospital organization, a local independent administrative institution.Correction: d, g this report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon ruptured of the foley catheter and came out.
|
| |
|
Search Alerts/Recalls
|