MEDTRONIC PUERTO RICO VILLALBA CAPSURE SP NOVUS LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 5592-53
Medical Device Problem Code Under-Sensing (1661)
Health Effect - Clinical Code Syncope/Fainting (4411)
Date of Event 10/11/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that the patient experienced syncope.It was also reported that the right atrial (ra) lead exhibited undersensing on stored electrograms (egm).It was also reported that the right ventricular (rv) lead had triggered an alert for rising impedance and exhibited oversensing on stored egms resulting in inappropriate pacing.The ra lead remains in use.The rv lead was capped and replaced. no further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPSURE SP NOVUS LEAD
Common Device Name
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23347981
Report Number2649622-2025-29533
Device Sequence Number20849095
Product Code DTB
UDI-Device Identifier00681490135733
UDI-Public(01)00681490135733(17)161013
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P850089
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/13/2016
Device Model Number5592-53
Device Catalogue Number5592-53
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Fifteen-Day
Initial Date Received by Manufacturer 10/12/2025
Initial Report FDA Received Date10/21/2025
Date Device Manufactured11/11/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5092-58 LEAD, ADDR01 IPG
Outcome Attributed to Adverse Event Required Intervention; Life Threatening; Hospitalization;
Patient Age81 YR
Patient SexMale
Patient Weight77 KG
-
-