• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 10MM; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 10MM; DENTAL IMPLANT Back to Search Results
Catalog Number TSVH10
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).G4: additional pma/510(k) number ¿ k011028 | k013227.It is unknown if the device will be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth # 30 fractured at the crest the doctor does not know yet what course of action will be taken.No additional information was provided.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation codes were added: 4109, 4111 and 3331.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H10: narrative/data was updated.Zimvie did not receive the reported implant for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Functional testing to recreate the reported event could not be performed due to the nature of the device & event.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.Based on the investigation and risk management file review, the most likely root causes determined from the investigation are parafunctional habits/patient factors, or customer error.Therefore, based on the available information, device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPL TAPERED SCR-V HA 3.7 MM 3.5MM 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos  gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23348500
Report Number0001038806-2025-02581
Device Sequence Number18527910
Product Code DZE
UDI-Device Identifier00889024019560
UDI-Public(01)00889024019560(17)210630(10)63338743
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K061410
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date06/30/2021
Device Catalogue NumberTSVH10
Device Lot Number63338743
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date10/21/2025
Supplement Report FDA Received Date11/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-