| Model Number |
610-2231011-01 |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation evaluated the device and the customer's report was not replicated or confirmed.The device was put through extensive testing, which included all functional testing using standard test accessories without duplicating the customer's report.No accessories were returned for evaluation.The device was recertified and returned to the customer.No trend is associated with reports of this type.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device was unable to obtain an ecg signal via the multifunction cable.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: this mdr was submitted after a retrospective review of complaints associated with capa-183.The submission was late because the mdr reports were incorrectly categorized as "not reportable" due to a misapplication of reporting criteria.This application stemmed from training, lack of clarity in procedural guidance, and inconsistent use of decision-making tools to ensure accurate classification.
|
| |
|