| Model Number |
WB991036 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that the olympus esg-100, displayed error code 246.The issue occurred during service or maintenance.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, d10, h6, h11.A review of the device history record found no deviations that could have caused or contributed to the reported issue.All records indicate that the product was manufactured, tested in accordance with all applicable procedures, and met all final product release criteria.The device was returned to olympus for inspection, and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the error e246 was due to a defective mother board.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No new additional information received from the customer.
|
| |
|