• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH BERLIN FACILITY OLYMPUS ESG-100, 220...240 V~,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH BERLIN FACILITY OLYMPUS ESG-100, 220...240 V~, Back to Search Results
Model Number WB991036
Medical Device Problem Code Communication or Transmission Problem (2896)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that the olympus esg-100, displayed error code 246.The issue occurred during service or maintenance.There were no reports of patient harm.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, d10, h6, h11.A review of the device history record found no deviations that could have caused or contributed to the reported issue.All records indicate that the product was manufactured, tested in accordance with all applicable procedures, and met all final product release criteria.The device was returned to olympus for inspection, and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the error e246 was due to a defective mother board.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No new additional information received from the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OLYMPUS ESG-100, 220...240 V~,
Common Device Name
OLYMPUS ESG-100
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH BERLIN FACILITY
rheinstrasse 8
teltow, brandenburg 14513
GM  14513
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH BERLIN FACILITY
rheinstrasse 8
teltow, brandenburg 14513
GM   14513
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23349294
Report Number3003724334-2025-00524
Device Sequence Number7770995
Product Code GEI
UDI-Device Identifier04042768910364
UDI-Public04042768910364
Combination Product (Y/N)N
PMA/510(K) Number
K073207
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 10/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWB991036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/24/2025
Supplement Date Received by Manufacturer10/30/2025
Initial Report FDA Received Date10/21/2025
Supplement Report FDA Received Date10/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/12/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-